Skip to main navigation Skip to search Skip to main content

Utilization of three-layers heterogeneous mammographic phantom through MCNPX code for breast and chest radiation dose levels at different diagnostic X-ray energies: A Monte Carlo simulation study

  • Princess Nourah Bint Abdulrahman University
  • Osmangazi University
  • Ural Federal University
  • Al-Azhar University
  • Dunarea de Jos University of Galati
  • Istinye University

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Introduction: We report the breast and chest radiation dose assessment for mammographic examinations using a three-layer heterogeneous breast phantom through the MCNPX Monte Carlo code. Methods: A three-layer heterogeneous phantom along with compression plates and X-ray source are modeled. The validation of the simulation code is obtained using the data of AAPM TG-195 report. Deposited energy amount as a function of increasing source energy is calculated over a wide energy range. The behavioral changes in X-ray absorption as well as transmission are examined using the F6 Tally Mesh extension of MCNPX code. Moreover, deposited energy amount is calculated for modeled body phantom in the same energy range. Results and discussions: The diverse distribution of glands has a significant impact on the quantity of energy received by the various breast layers. In layers with a low glandular ratio, low-energy primary X-ray penetrability is highest. In response to an increase in energy, the absorption in layers with a low glandular ratio decreased. This results in the X-rays releasing their energy in the bottom layers. Additionally, the increase in energy increases the quantity of energy absorbed by the tissues around the breast.

Original languageEnglish
Article number1136864
JournalFrontiers in Public Health
Volume11
DOIs
StatePublished - 2023

Keywords

  • MCNPX
  • Monte Carlo (MC)
  • X-ray
  • breast dosimetry
  • mammography

Fingerprint

Dive into the research topics of 'Utilization of three-layers heterogeneous mammographic phantom through MCNPX code for breast and chest radiation dose levels at different diagnostic X-ray energies: A Monte Carlo simulation study'. Together they form a unique fingerprint.

Cite this