Skip to main navigation Skip to search Skip to main content

Unexpected Artifact on 99mTc-Methylene Diphosphonate Bone Scintigraphy

  • University of Gezira
  • Fatima College of Health Sciences

Research output: Contribution to journalArticlepeer-review

Abstract

Asymmetric hot spots in the axial skeleton on bone scintigraphy may confound diagnosis. We describe an unexpected artifact of 99mTc-methylene diphosphonate near the breast in a 55-y-old woman with breast cancer. The initial whole-body bone scintigraphy revealed a solitary focal lesion in the anterior ribs on the left side. After careful tracking, we determined that this hot spot originated from the adhesive bandage. The patient had placed it in her left front pocket after removing it from the injection site. She was rescanned after the bandage had been removed from her pocket.

Original languageEnglish
Pages (from-to)360-361
Number of pages2
JournalJournal of Nuclear Medicine Technology
Volume52
Issue number4
DOIs
StatePublished - 1 Dec 2024

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • bone scintigraphy
  • breast cancer
  • unexpected artifact