T1 - Simulation of a Non-Contact Triple-Wavelength Reflectance Pulse Oximeter to Measure Oxygen and Carbon Monoxide Saturations
AU - Veerabathraswamy, Swathi
AU - John, Pauline
AU - Vasa, Nilesh J.
N1 - Publisher Copyright:
© 2023 IEEE.
PY - 2023
Y1 - 2023
N2 - This paper relates to the design of a novel reflectance pulse oximeter, which can measure not only the arterial oxygen saturation, but also that of carbon monoxide, in a non-contact manner. Such an instrument is especially relevant in the monitoring of patients who have a history of smoking, or those who have been exposed to carbon monoxide poisoning, such as from fires. The remote manner of measurement ensures that the risk of contact contamination is eliminated, and that the instrument can be used on infants as well, especially those born prematurely and having sensitive skin. A novel algorithm was devised which uses light sources of three different wavelengths to calculate the above mentioned saturations. A thorough sensitivity analysis was performed to determine the most suitable wavelengths for the purpose, which were determined to be 660 nm, 640 nm, and 940 nm. A system was designed to leverage the above mentioned algorithm, and the same was simulated using Simulink. The simulation gave satisfactory results, with maximum errors in calculation of oxygen and carbon monoxide saturations being 0.36% and 0.66%, respectively. These results are discussed along with measures that can be used to improve accuracy.
AB - This paper relates to the design of a novel reflectance pulse oximeter, which can measure not only the arterial oxygen saturation, but also that of carbon monoxide, in a non-contact manner. Such an instrument is especially relevant in the monitoring of patients who have a history of smoking, or those who have been exposed to carbon monoxide poisoning, such as from fires. The remote manner of measurement ensures that the risk of contact contamination is eliminated, and that the instrument can be used on infants as well, especially those born prematurely and having sensitive skin. A novel algorithm was devised which uses light sources of three different wavelengths to calculate the above mentioned saturations. A thorough sensitivity analysis was performed to determine the most suitable wavelengths for the purpose, which were determined to be 660 nm, 640 nm, and 940 nm. A system was designed to leverage the above mentioned algorithm, and the same was simulated using Simulink. The simulation gave satisfactory results, with maximum errors in calculation of oxygen and carbon monoxide saturations being 0.36% and 0.66%, respectively. These results are discussed along with measures that can be used to improve accuracy.
KW - Arterial blood
KW - carboxyhemoglobin
KW - deoxyhemoglobin
KW - oxyhemoglobin
KW - sensitivity
UR - https://www.scopus.com/pages/publications/85166482020
U2 - 10.1109/ICBSII58188.2023.10181063
DO - 10.1109/ICBSII58188.2023.10181063
M3 - Conference contribution
AN - SCOPUS:85166482020
T3 - Proceedings of the 9th International Conference on Biosignals, Images, and Instrumentation, ICBSII 2023
BT - Proceedings of the 9th International Conference on Biosignals, Images, and Instrumentation, ICBSII 2023
A2 - Kavitha, A.
A2 - Divya, B.
A2 - Karthick, S. Arun
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 9th International Conference on Biosignals, Images, and Instrumentation, ICBSII 2023
Y2 - 16 March 2023 through 17 March 2023
ER -