Abstract
We sought to determine effect of signal strength on mean retinal nerve fibre layer (RNFL) using Spectralis optical coherence tomography (S-OCT). Thirty normal subjects (18 female, mean 37.9 years, range 24-61) were imaged with S-OCT using variably dense Bangerter foils to alter Q value (1 unit signal strength = 4 units Q). We found a statistically significant (p < 0.01) linear relationship (R = 0.8643) between Q and RNFL (1 unit decrease Q = 0.181 um mean RNFL increase). Unlike previous observations of Cirrus and Stratus OCT, we found RNFL thickness does not decrease with decreasing signal strength in S-OCT.