Abstract
Patient-ventilator dyssynchrony is a mismatch between the patient’s respiratory efforts and mechanical ventilator delivery. Dyssynchrony can occur at any phase throughout the respiratory cycle. There are different types of dyssynchrony with different mechanisms and different potential management: trigger dyssynchrony (infective efforts, auto-triggering, and double triggering); flow dyssynchrony, which happens during the inspiratory phase; and cycling dyssynchrony (premature cycling and delayed cycling). Dyssynchrony has been associated with patient outcomes. Thus, it is important to recognize and address these dyssynchronies at the bedside. Patient-ventilator dyssynchrony can be detected by carefully scrutinizing the airway pressure-time and flow-time waveforms displaying on the ventilator screens along with assessing the patient’s comfort. Clinicians need to know how to depict these dyssynchronies at the bedside. This review aims to define the different types of dyssynchrony and then discuss the evidence for their relationship with patient outcomes and address their potential management.
| Original language | English |
|---|---|
| Title of host publication | Personalized Mechanical Ventilation |
| Subtitle of host publication | Improving Quality of Care |
| Publisher | Springer International Publishing |
| Pages | 269-285 |
| Number of pages | 17 |
| ISBN (Electronic) | 9783031141386 |
| ISBN (Print) | 9783031141379 |
| DOIs | |
| State | Published - 1 Jan 2022 |
Keywords
- Airway pressure-time waveform
- Assisted ventilator mode
- Auto-triggering
- Cycling dyssynchrony
- Double-triggering
- Dyssynchrony
- Flow dyssynchrony
- Flow-time waveform
- Ineffective efforts
- Patient-ventilator interaction
- Reverse triggering